| |
 |
 |
The ECI-1 monitors five key factors in corrosion –
linear polarization resistance, open circuit potential,
resistivity, chloride ion concentration, and temperature.
Toward that end, it includes multiple integrated electrodes,
directed by an on-board microcontroller, in a molded
plastic enclosure. Within a structure, ECI monitors
are connected to a multi-drop serial communications
network for power and data transmission. |
|
|
Linear Polarization Resistance
(LPR) and Open Circuit Potential (OCP): The ECI-1
measures linear polarization resistance by using a steel
working electrode, stainless steel counter electrode,
and manganese dioxide reference electrode. The working
electrode is a sacrificial component made of black steel,
designed to corrode at the same rate as the ASTM 615/A-compliant
steel that it monitors. Defective areas in protective
coatings over structural steel, such as epoxy or stainless
steel cladding, may be expected to demonstrate corrosion
characteristics comparable with those of black steel.
The ECI control module initiates the measurement of
open circuit potential between the working and reference
electrodes in the potentiostat circuit, and applies
the appropriate potentiostat drive potential between
the counter and working electrodes. A zero resistance
ammeter in the potentiostat circuit measures the cell
current.
The ECI scans cell current and drive potential over
a range about the OCP, and uses the resulting data to
calculate polarization resistance. The corrosion rate
of reinforcement steel may be expected to be inversely
proportional to this figure. If LPR is high and OCP
remains small in magnitude, managers may anticipate
that the reinforcement steel in a structure is passive,
suffering corrosion at a relatively low rate. As steel
begins to depassivate, due to an increase in chloride
ion concentration or other corrosive environmental conditions,
LPR will decrease and OCP will become increasingly negative.
Resistivity: The ECI-1 uses four stainless steel
electrodes to measure resistivity in the concrete that
surrounds it. A galvanostat circuit drives a stepped
current through the outer pair of these electrodes,
and measures the potential between the inner pair at
each step. Electronics within the ECI then perform a
linear regression to calculate the resistance between
the inner pair of electrodes. The ECI multiplies this
figure by the cell constant of its resistivity sensor
to derive the resistivity of the concrete in units of
ohms-cm. This provides information on the relative amount
of moisture in the concrete. Structure managers may
also use this resistivity parameter with the geometric
cell constant of the working, counter, and reference
electrodes to correct for ohmic resistance errors in
polarization resistance measurements.
Chloride Ion Concentration: The ECI-1 uses a
silver/silver-chloride ion specific electrode in combination
with its reference electrode to measure chloride ion
concentration. Over time, a potential will develop between
the Ag / AgCl and reference electrodes. The magnitude
of this potential is related to the chloride concentration
in the concrete surrounding the instrument. The ECI
reports chloride measurement results as a potential.
Temperature: The ECI-1 includes an on-board solid
state sensor, which provides information on the temperature
within the concrete surrounding it.
Microcontroller: An on-board microcontroller
sequences all ECI sensor measurements. It also controls
sensor drives and data acquisition through digital-to-analog
and analog-to-digital converters, and performs all necessary
calculations for corrosion measurements. The microcontroller
provides power management for the ECI in low power remote
installations, such as battery- or solar-powered operations.
Toward that end, it can conserve energy by placing various
system components into low-power or off modes.
Enclosure: The ECI-1 enclosure is engineered
to provide environmental and structural protection for
its sensors and electronics without compromising the
integrity of the structure in which it is embedded.
This molded plastic enclosure protects the instrument’s
electronics against moisture and chemical encroachment,
while providing a rigid base for its electrodes. A flexible
waterproof and chemically resistant potting compound
is used within the ECI to provide further protection,
and to cushion the electronics from external stress
on the enclosure. During installation within a structure,
a small cage of #3 rebar may be placed around the ECI
to further isolate the instrument from mechanical stresses.
Such a cage, attached to the structure’s reinforcement
mat, may also serve to hold the ECI at an appropriate
level for long-term monitoring.
Communications and Network Management: Within
a structure, ECI instruments are connected to a multi-drop
serial communications network. Various network configurations
and protocols may be used. The preferred implementation,
however, is a local area network using the SDI-12 protocol.
Under this protocol, each ECI is connected to a datalogger
interface using a three-wire sensor operating at a 1,200
bps data rate. The SDI-12 bus consists of +12 volts,
ground, and data lines. Each sensor on the bus is individually
addressable. Many dataloggers and sensor manufacturers
support the SDI-12 protocol.
Within an ECI installation, the data collection system
– the datalogger – is located external to
the structure under observation, within an environmentally
protected enclosure, such as a NEMA-4 box. The datalogger
connects to the multi-drop serial communications network
cables exiting the structure. It supplies power to the
SDI-12 network and to all connected instruments. The
datalogger itself may be powered by local electric lines
or by a battery recharged by a solar collector. It may
be programmed to switch ECI instruments on and off periodically,
and to issue commands to collect and transmit data.
The datalogger can timestamp all corrosion measurements,
and specify the identification number and location of
the instrument providing each. It can download these
data to a portable computer on-site, or transmit them
via wireless transceiver or cellular modem for remote
data collection and operation.
|
|
|
|
Application: The ECI-1 is
used to monitor the corrosion of reinforcement steel
in concrete structures, including bridges, buildings,
dams, erosion control structures, flood control channels,
parking garages, piers, pylons, roadways, and spillways.
By way of example, in bridge decks, ECI corrosion monitors
are installed within the bridge during constructrion
– before the concrete is poured. The instruments
are placed with their electrodes facing the top surface
of the bridge deck, at the level of the top layer of
reinforcement steel. This orientation ensures that the
ECI’s sensor electrodes encounter the same environmental
and corrosion conditions as the reinforcement steel
they are monitoring.
Once installed, the ECI instruments generate readings
on LPR, OCP, resistivity, chloride ion concentration,
and temperature within reinforced concrete. These
readings can be processed to form a “corrosion
map” of a structure. Managers can use this information
to determine the location, timing, and type of maintenance
that are needed by the structure. By monitoring the
corrosive environment within a structure, managers
can forecast its remaining life and schedule its remediation
or replacement without using costly, time-consuming,
and labor-intensive destructive evaluation methods.
|
|
|
|
 |
 |
Physical Dimensions
Enclosure Material
Chloride Voltage
Trend Indicator
Conductivity/Resistivity Measurement
Polarization Resistance Measurement
Temperature Sensor
Open Circuit Potential
Estimated Power Requirements
Communications
back to top > |
Enclosure & Electrodes:
83mm (L) x 94mm (W) x 122mm (H)
VALOX™ Plastic, Epoxy Potted, Water Tight Seal
Range: +/- 1.3 Volts
Electrodes(2): Ag/AgCl 15mm (L) x 1mm (Dia.), MnO2
reference electrode, Force Institute Model ERE 20
Range: 1,000 to 19,000 Ohm-cm
Electrodes(4): 316L SS (4) 12mm (L) x 1mm (Dia.) spaced
at 8mm
Range: 1 KOhm-cm2 -> 1 MOhm-cm2
Electrodes(3): 316L SS counter electrode (1) 18cm2
x 1mm thick, MnO2 reference electrode, Force
Institute Model ERE 20, Steel working electrode 15.5mm
(Dia.) x 10.0mm (H)
Range: -40° C to +70° C
Range: +/- 1.3 Volts
Inactive: 1.5 mAmps @ 12 Volts
Active: 4.5 mAmps @ 12 Volts
SDI-12 V1.2 compatible |
|
|
 |
|