Medical Interface: Q-Stim’s
medical interface is a PC software application that
makes it easy for healthcare providers to set stimulation
parameters. From an intuitive graphical user interface,
clinicians key in the dimensions of pulse amplitude,
duration, and frequency. They can set these controls
for all pacing channels at once, or for each independently.
They can also set the sequence and timing among channels,
establishing the optimal flow of electrical pulses
across the four points of the Q-Stim pacing lead.
Telemetry Unit: Once a patient’s stimulation
profile is complete, clinicians upload it to Q-Stim’s
telemetry unit through a wired or wireless connection.
The telemetry unit encodes all pacing parameters,
and transmits them transcutaneously via low-frequency
carrier signal to the stimulation generator.
Stimulation Generator: Based on the prescribed
parameters, the stimulation generator calculates the
charge needed at each channel of the pacing lead.
It periodically tests the interface between this lead
and the patient, obtaining a real-time electrical
characterization, including measures of impedance
and of the rate at which energy dissipates into the
tissue. Based on these tests, and on the stimulation
profile, the generator sends a precisely measured
charge to a capacitor via low-power dissipation. The
capacitor, in turn, passes that charge to the lead,
out of circuit from the pulse generator. By adapting
its electrical charge to conditions at the electrode-tissue
interface – the connection between the pacing
lead and the patient – Q-Stim intelligently
compensates for aging of the system and changes in
the patient’s tissue condition.
With this adaptive technology, and with capacitive
coupling to separate pulse generation from delivery,
Q-Stim ensures that only the charge needed for stimulation
is sent to the patient. In this way, it promotes patient
safety, guarding against over-stimulation. It also
maximizes energy efficiency, revolutionizing the state
of the art in pacemaker technology. Older pacemakers
could not adapt their stimulation on the basis of
frequent and complete electrical characterization
of the tissue interface, and so generated excess charge
to prepare for a worst-case scenario. This, however,
wasted energy and shortened battery life. Q-Stim,
by contrast, uses its proprietary adaptive technology
to maximize energy efficiency, and to conserve the
lives of both the system battery and the pacing lead.
That, in turn, promotes patient safety by reducing
the need to replace pacemakers over time.
Multi-Channel Pacing Lead: The final component
of the Q-Stim system is a multi-point pacing lead,
which delivers stimulation to the patient through
any or all of four channels, in any sequence. By stimulating
at multiple points, progressing from one to the next,
Q-Stim can supplement or entrain the natural electrical
flow that governs many functions of the body. In the
stomach, for example, Q-Stim can progressively pace
from proximal to distal regions, helping regulate
the gastric slow wave and accelerate the flow of nutrients
in patients who suffer gastroparesis, or delayed gastric
emptying. By reversing this sequence – that
is, by pacing from the distal to the proximal stomach
– Q-Stim can promote feelings of fullness or
satiety in patients who suffer obesity, and so help
reduce caloric intake.
Virginia Technologies has developed
reference designs and a prototype neurostimulator,
demonstrating the Q-Stim technology. The company would
be happy to review these with medical device developers
who wish to license this technology for their own
neurostimulation or pacemaker systems. To explore
this opportunity, please
click
here.
back to top >